Dimensions..................................................... 8 cm (3�) W x 15 cm (6�) H x 2.5 cm (1�) D
Weight............................................................ 280 g (10 oz) with alkaline batteries
Oxygen Saturation Display Range................ 0 � 100% SpO2
Pulse Rate Display Range.............................. 18 � 321 beats per minute (BPM)
Saturation Declared
Accuracy (Arms)*............ 70 � 100% SpO2
Adults/Pediatrics Neonates
No Motion: Finger Clip �2 digits N/A
Flex �3 digits N/A
Motion: Finger Clip �3 digits N/A
Flex �3 digits N/A
Low Perfusion: Finger Clip �3 digits N/A
Flex, Flexi-Form� II �3 digits �4 digits
Pulse Rate Declared Accuracy
No Motion................................................. 18 � 300 BPM
Motion....................................................... 40 � 240 BPM
Low Perfusion............................................ 40 � 240 BPM
Adults/Pediatrics Neonates
No Motion: Finger Clip �3 digits N/A
Flex, Flexi-Form II �3 digits �3 digits
Motion: Finger Clip �5 digits N/A
Flex �3 digits N/A
Low Perfusion: Finger Clip �3 digits N/A
Flex, Flexi-Form II �3 digits �3 digits
Measurement Wavelengths and Output Power**
Red............................................................ 660 nanometers @ 0.8 mW maximum average
Infrared...................................................... 910 nanometers @ 1.2 mW maximum average
Indicators
Pulse Quality Display................................... LED, tricolor
Numeric Displays........................................ 3-digit 7-segment LEDs, red
Temperature
Operating................................................... -20 to +50�C (-4 to +122�F)
Storage/Transportation............................... -30 to +50�C (-22 to +122�F)
Humidity
Operating................................................... 10 � 90% noncondensing
Storage/Transportation............................... 10 � 95% noncondensing
Altitude
Operating................................................... Up to 12,000 meters (40,000 feet)
Hyperbaric Pressure.................................... Up to 4 atmospheres
Power Requirements..................................... Six 1.5V AA-size alkaline batteries.
100 hours maximum display brightness;
160 hours with normal brightness.
Classifications per IEC 60601-1 / CSA601.1 / UL60601-1
Type of Protection:..................................... Internally powered (on battery power)
Degree of Protection:................................. Type BF-Applied Part
Mode of Operation:................................... Continuous
Enclosure Degree of Ingress Protection:...... IPX2
* � 1 Arms represents approximately 68% of measurements